| Model Number: | HP-EM4-UV 4 |
| MOQ: | 1 unit |
| Payment Terms: | T/T |
| Parameter | Specification |
|---|---|
| Spectral Ranges | UVA: 315nm–400nm (λₚ = 365nm) UVB: 280nm–315nm (λₚ = 310nm) UVC: 230nm–280nm (λₚ = 254nm) UVV: 395nm–445nm (λₚ = 420nm) |
| Compatible Light Sources | High-pressure mercury lamps, iron lamps, gallium lamps, metal halide lamps, xenon lamps, etc. |
| Sampling Speed | 2048 times/second (Typical) |
| Power Storage Interval | 32 times/second |
| Temperature Storage Interval | 2 times/second |
| Recording Cycle | 32 minutes |
| Measurement Accuracy | For H < 5 mW/cm²: ±0.5 mW/cm² For H ≥ 5 mW/cm²: ±10% H |
| Measurement Units | μW/cm², mW/cm², W/m² |
| Device Dimensions | Diameter: 120 mm |
| Sensor Position | Back of the display screen |
| Data Export | Supported via USB connection |